" TINTOMETER"
: Lovibond™ B.O.D. IR-Transmitter-Sensomat,
ET 500 & ET 510, Intelligent B.O.D. Measuring
System using IR Data Capture & Data Evaluation
Software, Lovibond™ 적외선 B.O.D. 측정
시스템 |
|
|
■ Up to 144 sets of results can be accessed with infrared
remote operation
■ Variable Measuring Time from 30 Minutes to 99 Days
■ User-friendly Means of BOD Measurement
■ Data Evaluation Software with Statistical Functions
■ Electronic storage ensures confidence in results
and other measurement data
■ Complies with GLP(Good Laboratory Practice)/ISO 9000
특징 (Features)
* Respirometric Determination of Biochemical Oxygen
Demand(B.O.D.)
* Organic substances in the sample are oxidized
microbially by the use of oxygen (O2).
The resultant carbon dioxide (CO2) is absorbed by
potassium hydroxide(KOH) and this results in a pressure
drop in the system which is proportional to the BOD value.
This is detected by the IR-Sensor, logged by
the BOD-Sensomat
and converted directly into mg/l of BOD.
* Intelligent Measuring Systems ET 500 & ET 510 구성 :
ㆍIR-Sensor
B.O.D. Bottle 위에 뚜껑 형태로 간단히 끼워지는 B.O.D.
측정 센서 Measurement Device인 동시에 Storage Unit 역할
Together with the Sample Bottle, it forms a Mobile,
Self-contained Measuring Unit.
ㆍBOD-Sensomat : Portable "ore”of the System
한번의 버튼 조작으로 모든 측정 데이터를 기록 및 조회
간단하고 장력한 Evaluation 및 Documentation 기능 제공
ㆍInductive Stirring-System
주기적인 Acceleration과 Stirring Rod의 Centering을 통하여
Water Sample과 Air Phase 사이의 최적화된 Gas-exchange를 보장
규격 (Specification)
(BOD-Sensomat)
* Parameter : BOD n (DIN 38 409-H 52)
* Measuring Ranges : Programmed : 0...40 / 80 / 200 / 400 /
800 / 2000 / 4000 ㎎/ℓ B.O.D
* Test Duration : 30 Minutes to 99 Days Measurement
* Intervals : 10 Seconds to 24 Hours
* Mesurement points : 1 to 144 Operable in Parallel
* Data Storage : 180 to 360 Data-Sets(Data, Time,
Measurement, Identification of sample
& measurement point)
* Data security : 100% fixed
* Memory display : LCD 128 ×64 Pixel / 6 rows
* Auto-Start Function : on/off switching facility,
measurement begins after pressure drops, or maximum
waiting period
* Software : Calculation of Mean, Standard Deviation,
exclusion of individual rogue samples
* Approval Marks : TUV/GS, CE, UL, CUL(UL/CSA)
* Degree of protection : IP42, DIN 40 050
* Batteries : 3 mignon/type AA, alkaline 1.5V,
1 Lithium CR 2430
* Ambient temperature : (Storage) -25 ... +65℃ /
(Operation)+5 ... +40℃
* Weight : Approx. 390g including Batteries
* Dimensions : L100 ×w200 ×h45mm
(IR-Sensor)
* Principle of Operation : Manometric/Respirometric with
Electronic Pressure, relative pressure measurement,
IR-transmitter. receiver ±1% of measured value ±1 hPa
* Range : -500/+350 hPa(=mbar)
* Resolution : ±1 hPa
* Data Storage : 180 to 360 Data-Sets(Data, Time,
Measurement, Identification of sample & measurement point)
* Data Security : 100% fixed
* Memory display : LED for recognition
* Batteries : 2 Lithium CR 2430(approx. 2~5 years)
* Approval Marks : TUV/GS, CE, UL, CUL(UL/CSA)
* Degree of Protection : IP 54 DIN 40 050
* Ambient Temperature : (Storage)-25℃ ... +65℃/
(Operation)+5 ... +40℃
* Weight : 90g including Batteries
* Dimension : Φ70 mm, 70 mm in Height
|
|
|
CAT No. |
Description |
단위/갯수 |
Price[VAT포함] |
|
|
|
|
|
|
|