::: " FLUKE" : FLUKE 71x SERISE Calibrator SPECIFCATION :::

       
       
 
 
새 페이지 1
 
" FLUKE" : FLUKE 71x SERISE  Calibrator   SPECIFCATION



71x Serise 712 713
714 715 716

Model

Parameters

Range

Resolution

Accuracy

Notes

.712 RTD

W

15W to 3200W

0.1W

0.1W  to 1W

Pt ;100 200 500 1000 (385)
Pt 100 (392)
Pt 100 (392) JIS
Ni 120 (672)

RTD

-200 to 800°C
(Pt 100)

0.1°C, 0.1°F

0.33°C, 0.6°F
(Pt 100)
.713 30G
.Pressure
.Over Pressure
.3xFS
Pressure 0-30 Psi
(207 kPa)
0.001 Psi
0.01 kPa)
0.05% FS Gasses/liquids(non-corrosive)
Zero, Min, Max, Hold, Damp


Units; Psi, in. H2o(4
°C)
H2o(20°C)
, kPa,cm H2o(4°C)
cm H2o (20°C), bar, mbar,
kg/cm2, mmHg, inHg
mA 0.001 mA 0.001 mA 0.025% +/-
1 count
.713 100G
.Pressure
.Over Pressure
.2xFS
Pressure 0-100 Psi
(690 kPa)
0.01 Psi
0.1 kPa)
0.05% FS
mA 0 to 24 mA 0.001 mA 0.025% +/-
1 count
714
Thermocouple
mV -10 to 75 mV 0.01 mV 0.025%+
1 count
9TC types; JKTERSB
per NIST 175 and ITS-90
L U per DIN 43710and IPTS-68




TC input
     ouput
-200 to 1800°C
depending on
type (K,-200
to 1370°C)
0.1°C or °F
(1°C or °F
BRS)
0.3°C+10 mV


715 Volt/mA mA 0 to 100 mV 0.1mV 0.025%+ FS Sink(simulate trasmitter);
Display mA or % of span;
24 volt loop supply
Drive; 20mA into 1000W for
Battery>6.8V; 700W for battery
5.8 to 6.8V
V 0 to 10V 0.001V 0.02%+2counts
mA/% of span 0 to 24 mA
(-25% to 125%)
0.001 mA 0.02%+2counts
Loop Supply 24 Vdc N/A +/- 10%
716 pressure pressure 10,000 Fs'g . 0.05%
mA 0 to 24mA 0.001mA 0.025%