새 페이지 1
" FLUKE"
: Fluke 787,789
PROCESSMETER SPECIFCATION |
|
|
|
|
787
|
789
|
Measurement Function
|
Best Accuracy Range
and Resolution
|
(% of reading + LSD)
|
V dc |
400.0㎷, 4.000 V, 40.00 V, 400.0 V, 1000 V |
0.1% + 1 |
V ac (true-rms) |
400.0㎷, 4.000 V, 40.00 V, 400.0 V, 1000 V |
0.7% + 2 |
㎃ dc |
30.000 ㎃ |
.05% + 2 |
A dc |
1.000 A (0.440 A continuous) |
0.2% + 2 |
A ac |
1.000 A (0.440 A continuous) |
1% + 2 |
Resistance |
400.0 Ohms, 4,000 k, 40.00 k, 400.0 k, 4.0
M, 40 M |
0.2% + 1 |
Frequency (0.5 ㎐ to 20 ㎑) |
199.99 ㎐, 1999.9 ㎐, 19.999
㎑ |
.005% + 1 |
Diode Test |
789:2.000V (shows diode voltage drop)
787:2.400V (shows diode voltage drop)
|
2 % + 1
2 % + 1 |
continuity |
Beeps for resistance < approx. 100 ohms |
|
Output Function
|
Range and Resolution
|
Drive Capability
|
Accuracy (% of span)
|
DC Current Output
(Internal battery operation)
|
0.000 to 20.000 ㎃ or 4.000 to 20.000
㎃ (selectable at power-up) Over-range to 24.000 ㎃ |
789:24 V compliance or, 1,200 Ohms, @ 20 ㎃
787:12V compliance or, 500 Ohms, @ 20 ㎃
|
,05%
,05% |
DC Current Simulate
(Ext. 24 Volt loop supply) |
0.000 to 20.000 ㎃ or 4.00 to 20.000 ㎃,
(selectable at power-up) Over-range to 24.000㎃
|
1000 Ohms,
@ 20㎃ |
,05% |
24 V Loop Supply |
789: Minimum 24V
787:not available
|
250 Ohms
@ 20㎃
|
>24V |
Current Adjustments Modes |
Manual:Coarse,Fine,25% and
100% step
Automatic:Show Ramp,Fast Ramp,25% step |
|
|
|
|
|
|